Skip to product information
1 of 1
PROMOTION FREE software bundles available on select oscilloscope and spectrum analyzer purchases through December 31, 2026.See Eligible Models →
New

MSO6000HDP-PWR

Regular price $1,480.00
Sale Contact Us For Lead Time

View full details

MSO6000HDP-PWR Power Analysis Option

Overview

The MSO6000HDP-PWR turns an MSO6000HDP Series oscilloscope into a power analyzer. Choose an analysis, connect voltage and current probes as the on-screen diagram shows, and the oscilloscope configures its channels and reports results as waveforms, a result table and statistics.

Twelve analyses cover a converter from the AC input to the output, with 12-bit vertical resolution for small ripple and switching details.

The option is enabled by a software license keyed to the oscilloscope's serial number and is permanent for the life of the instrument.

Key Features

  • Input analysis — power quality (RMS and peak voltage and current, crest factors, real, reactive and apparent power, power factor, phase angle), harmonics against IEC 61000-3-2 Class A, B, C or D, and inrush current.
  • Switching device analysis — switching loss, RDS(on), slew rate and safe operating area.
  • Output and loop analysis — ripple, modulation, turn-on/turn-off time, transient response and efficiency.
  • Lifetime license — activated once, permanent for the life of the instrument.

When You Need This Option

  • SMPS and converter design: switching loss, SOA and efficiency at every load point.
  • Power-integrity checks on low-voltage, high-current rails.

Compatibility

For the MSO6000HDP Series: MSO6204HDP and MSO6104HDP. Keep the oscilloscope on the latest UNI-T firmware.

Frequently Asked Questions

Do I need special probes?

Yes. Power analysis needs a current probe and, for floating or line-referenced measurements, a high-voltage differential probe. Current probes: the UT-P43 (25 MHz, 20 A) or UT-P44 (50 MHz, 40 A) for fast switching edges; the UT-P4150 (12 MHz, 150 A) for higher currents; the UT-P40 (60 A), UT-P41 (100 A) or UT-P42 (200 A) for line-frequency harmonics and efficiency. High-voltage differential probes: UT-P30 through UT-P36.

How is this option activated?

Buy the option from your UNI-T distributor with the oscilloscope's serial number. You receive a license certificate; register it on the License Activation page of the UNI-T website with the instrument serial number to get the license file. Copy the file to the root of a USB drive, connect it to the oscilloscope and install it from the Option Install menu. Utility > Options then shows the option as Activated, with a trial duration of Permanent.

Can I try it first, and is the license permanent?